Performance Quick Check

Verifying your Electrical Calibration

32
Passing this test ensures that the electrical calibration is within
expected performance.
1. Follow the steps in the sections "Preparation " on page 30
and "Network analyzer settings and electrical calibration" on
page 30.
2. Connect the 85054-60037 "type-N female to female adapter"
(provided with the 85092C option 00A ECAL kit) between the
two 8120-8862 "type-N test port cables" supplied with the
LCA.
3. Press the [Macro Setup] key on the front panel of the network
analyszer, then select
4. Click
to start the LCA measurement setup.
5. Select the EE Measurement tab.
6. Click
.
7. Make a corrected measurement with the network analyzer.
8. Inspect the results.
• Electrical return loss S11 should be similar to S22.
• Both curves should be below -20dB over the full
measurement range, with only low frequency dependence
on applied modulation frequency.
• Transmission loss S12 should be similar to S21.
• Both curves should be within +0.1dB to -0.6dB over the
full measurement range, with only low frequency
dependence on applied modulation frequency.
If the results differ from the expected values
• Visually inspect the electrical connectors on the "Electronic
Calibration Modules" and the .
If an electrical connector seems to be damaged, contact
Keysight for repair or replacement.
• Clean the electrical connectors on the "Electronic Calibration
Modules" and the "Flexible Cables for Test Ports.
If an electrical connector seems to be damaged, contact
Keysight for repair or replacement.
Keysight N4374B Lightwave Component Analyzer, User's Guide