Tornado 2+
EN
Installation and operating instructions
0297
5286100072L02
*5286100072L02*

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Tornado 2+ and is the answer not in the manual?

Questions and answers

Summary of Contents for Durr Dental Tornado 2+

  • Page 1 Tornado 2+ Installation and operating instructions 0297 5286100072L02 *5286100072L02*...
  • Page 3 Contents Contents Information about electrical con- nections ....Transport ......Important information Installation .
  • Page 4 Contents Troubleshooting 15 Tips for operators and service techni- cians ......Appendix 16 Handover record ....5286100072L02 1909V005...
  • Page 5: About This Document

    Important information The signal word differentiates between four levels Important information of danger: – DANGER Immediate danger of severe injury or death About this document – WARNING Possible danger of severe injury or death These installation and operating instructions rep- resent part of the unit.
  • Page 6 Important information Copyright information Safety All circuits, processes, names, software pro- Dürr Dental has designed and constructed this grams and units mentioned in this document are unit so that when used properly and for the protected by copyright. intended purpose it does not pose any danger to The Installation and Operating Instructions must people or property.
  • Page 7: General Safety Information

    Important information Electrical safety WARNING Observe and comply with all the relevant elec- ❯ Risk of explosion due to ignition of trical safety regulations when working on the combustible materials unit. Do not operate the unit in any rooms in ❯...
  • Page 8 Important information 2.10 Disposal Unit The unit must be disposed of properly. Within the European Union, the unit must be disposed of in accordance with EU Directive 2012/19/EU (WEEE). If you have any questions about the correct ❯ disposal of parts, please contact your dental trade supplier.
  • Page 9: Product Description

    Product description Product description Overview Scope of delivery The following items are included in the scope of delivery (possible variations due to country-spe- cific requirements and/or import regulations): Tornado 2+ ....5286100036. –...
  • Page 10 Product description Technical data Electrical data 5286100036 Nominal voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Speed 1390 1650 Type of protection IP20 Mains fusing * Max. permissible mains impedance in accordance with EN 61000‑3‑11 0.2069 Circuit breaker fuse characteristics C in acc.
  • Page 11 Product description Network connection LAN technology Ethernet Standard IEEE 802.3u Data rate Mbit/s Connector RJ45 Type of connection Auto MDI-X Cable type ³ CAT5 Ambient conditions during storage and transport Temperature °C -10 to +55 Relative humidity max. 95 Ambient conditions during operation Temperature °C +10 to +40...
  • Page 12: Type Plate

    Product description Distance between rubber feet Compressor unit type plate Membrane drying unit The type plate of the membrane drying unit is located on the membrane drying unit. (cm) (cm) (cm) (cm) (cm) 20 l 17.5 Type plate Complete system The type plate for the complete system is located on the pressure tank.
  • Page 13: Operation

    Product description Operation Compressor unit Air intake filter Safety valve Condensate drain valve Pressure tank Pressure gauge/display Compressed air connection (quick release coupling) Pressure relief solenoid valve Non-return valve 10 Condensate collector tray 11 Automatic/manual condensate drain valve, membrane drying unit 12 Sintered filter for membrane drying unit 13 Membrane drying unit 14 Electronic control with displays and operating elements...
  • Page 14: Operating Panel

    Product description All of the measurement data for the appliance comes together in the controller (e.g. pressure in the pressure tank), where it is then evaluated. Likewise, various settings (e.g. switch-on/cut-off pressure) can be adjusted, or the unit can be connected via the network to Tyscor Pulse. Operating panel Pressure range display/adjustment Service key with blue LED...
  • Page 15: Installation/Setup Room

    Assembly Assembly ≥ 40 °C Requirements The unit must not be set up or operated within the vicinity of the patients (within a radius of 1.5 m). The unit can be installed either at the same level as the surgery room or on a floor below (e.g. cel- lar).
  • Page 16: Installation

    Assembly Transport Installation Establishing the compressed WARNING air connection Risk of explosion of the pressure tank and pressure hoses The supplied flexible pressure hose The pressure tank and the pressure ❯ between the pipe system and the com- hoses must be vented before they are pressor prevents vibrations from being stored or transported.
  • Page 17: Network Connection

    Assembly drained condensation, it is collected in the collec- tor tray. As an option, the condensation can also be drained through a hose into the waste water system. Always comply with appli- cable national regulations for waste water systems. Place a collector tray under the condensate ❯...
  • Page 18 Assembly Plug the male connector of the mains connec- Connecting pressure vessels ❯ tion into the female mains connector on the If two devices are connected to a single com- controller. pressed air network, pressure equalisation must take place between the pressure vessels. To do The male mains connector is mechani- this, the pressure vessels need to be connected cally secured so that it cannot drop out.
  • Page 19: Checking The Safety Valve

    Assembly Commissioning Before taking the unit into operation, check for ❯ any damage. Damaged units must not be used. In many countries technical medical prod- Main Main ucts and electrical devices are subject to regular checks at set intervals. The owner must be instructed accordingly.
  • Page 20 Assembly pressure tank. The condensation water can only DANGER be drained from the pressurised pressure tank. Start up the unit and wait until the cut-off pres- Risk of explosion of the pressure tank ❯ sure is reached. and pressure hoses Do not change the safety valve set- Pressure tank ❯...
  • Page 21: Adjustment Options

    Assembly 10 Adjustment options Combining devices safely – The overall safety of the unit and its main per- 10.1 Adjustment of the switch- formance features are independent of the net- work. The device is designed for operation on/cut off pressure independent of a network.
  • Page 22: Circuit Diagram

    Assembly 11 Controller 11.1 Circuit diagram X100 X100 Main PCB (main board) Fuse for motor, valve, fan Fuse for controller Supply voltage, 230 V Compressor unit Relief valve Cooling fan motor, membrane drying unit Network and CAN bus Connector to the display PCB Display/controller PCB X100 Connector to the main PCB...
  • Page 23 Assembly 11.2 Electrical connections, controller H114 H112 H109 H108 H107 H110 X100 H113 H111 H100 H103 H101 H102 MAIN µSD DOWN: CAN UP: ETHERNET Main PCB (main board) Fuse for motor, valve, fan Fuse for controller Supply voltage, 230 V Compressor unit Relief valve Cooling fan motor, membrane drying unit...
  • Page 24 Usage Pressure range Usage The pressure is displayed and can be adjusted in this area. The pressure is displayed via: 12 Operation 1. Blue LED (£ 4.5 bar): only illuminates while the pressure is building up Prior to working on the unit or in case of during start-up operation danger, disconnect it from the mains.
  • Page 25 Usage This keeps the operator informed about emer- gency mode, humidity, leaks or overheating. Warning messages and information are automati- cally deactivated after the fault has been rectified, with exception of emergency mode. 5286100072L02 1909V005...
  • Page 26: Maintenance Schedule

    Usage 13 Maintenance Prior to working on the unit or in case of danger, disconnect it from the mains. CAUTION Risk of infection due to burst filters Particles enter the compressed air network and can therefore enter the mouth of the patient. Replace filters in accordance with the maintenance schedule.
  • Page 27 Usage Remove the air intake filter. ❯ 13.2 Replacing the air intake filter Switch the compressor to standby mode with ❯ the standby button. Unplug the mains plug. ❯ Take off the noise reducing hood and the foam ❯ filter cover. Insert a new air intake filter.
  • Page 28: Taking Out Of Use

    Usage 14 Taking out of use Sintered filter Unscrew and remove the filter housing. ❯ 14.1 Taking the unit out of use Remove the sintered filter. ❯ Insert a new sintered filter. If the unit is not to be used for a prolonged ❯...
  • Page 29: Storage Of The Unit

    Usage 14.2 Storage of the unit WARNING Risk of explosion of the pressure tank and pressure hoses The pressure tank and the pressure ❯ hoses must be vented before they are stored or transported. Protect the unit against moisture, dirt and ❯...
  • Page 30: Troubleshooting

    Troubleshooting Troubleshooting 15 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. Prior to working on the unit or in case of danger, disconnect it from the mains. Error Possible cause Remedy...
  • Page 31 Troubleshooting Error Possible cause Remedy Water dripping from air con- Maintenance work not carried Regularly drain the condensa- ❯ out regularly (without membrane tion water from the pressure sumers drying unit) tank, see "9.3 Draining the condensation water" Defective membrane drying unit Inform a service technician.
  • Page 32 Appendix Appendix 16 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)
  • Page 36 Hersteller/Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...