Table of Contents

Advertisement

Primo, Duo, Trio, Quattro, Quattro
Tandem
EN
Montage- und Gebrauchsanweisung
0297

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Primo and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Durr Dental Primo

  • Page 1 Primo, Duo, Trio, Quattro, Quattro Tandem Montage- und Gebrauchsanweisung 0297...
  • Page 2 The current version of the installation and operating instructions is available in the Download Center: http://qr.duerrdental.com/5152100003 © Dürr Dental SE...
  • Page 3: Table Of Contents

    Technical data ....11.1 Adjusting the pressure switch . . . Primo ..... 11.2 Adjusting the motor protection Duo .
  • Page 4 Contents 14 Maintenance ..... 14.1 Maintenance schedule ..30 14.2 Wear parts and replacement parts .
  • Page 5: Important Information

    – NOTICE These installation and operating instructions Risk of extensive material/property damage apply to: Primo Other symbols REF: 5152-01 These symbols are used in the document and on or in the unit: REF: 5252100001; 5252-01; 5252-01/Service;...
  • Page 6: Copyright Information

    Important information Safety Manufacturer The unit has been developed and designed in such a way that dangers are effectively ruled out if used in accordance with the Intended Use. Copyright information Despite this, the following residual risks can All circuits, processes, names, software pro- remain: grams and units mentioned in this document are –...
  • Page 7: General Safety Information

    Important information Electrical safety WARNING Observe and comply with all the relevant elec- ❯ Risk of explosion due to ignition of trical safety regulations when working on the combustible materials unit. Do not operate the unit in any rooms in ❯...
  • Page 8: Disposal

    Important information 2.10 Disposal The unit must be disposed of properly. Within the European Union, the unit must be disposed of in accordance with EU Directive 2012/19/EU (WEEE). If you have any questions about the correct ❯ disposal of parts, please contact your dental trade supplier.
  • Page 9: Product Description

    Duo Tandem, Trio and Primo 230 V, 1~, with membrane dry- Quattro Compressors ... 4251-500-00 ing unit ......5152-01...
  • Page 10: Technical Data

    Product description Technical data Primo Electrical data 5152-01 Nominal voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection switch, recommended setting Type of protection IP 24 Mains fuse* Circuit breaker fuse characteristics B, C or D in accordance with EN 60898-1...
  • Page 11: Duo

    Product description Electrical data 5252-01 5252-51 5252-01/Service 5252100024 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection switch, recommended settings Type of protection IP 24 IP 24 Mains fuses * Max. permissible mains impedance in £ 0.142 accordance with EN 61000‑3‑11 ** Circuit breaker fuse characteristics B, C or D in acc.
  • Page 12 Product description Classification Medical Device Class (MDR) 5152100003L02 2305V004...
  • Page 13: Duo

    Product description Electrical data 5252100001 Nominal voltage 110-115 110-127 Mains frequency Nominal current at 8 bar (0.8 MPa) 14.1-14.6 14.2-13.3 Motor protection switch, recommended setting Type of protection IP 24 Mains fuse* Circuit breaker fuse characteristics B, C or D in accordance with EN 60898-1 General technical data Pressure tank volume Suction power, approx.
  • Page 14: Trio

    Product description Trio Electrical data 5352-01 5352100001 5352100013 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection switch, recommended settings Type of protection IP 24 Mains fuses * Max. permissible mains impedance in £ 0.078 accordance with EN 61000‑3‑11 Circuit breaker fuse characteristics B, C or D in acc.
  • Page 15 Product description Classification Medical Device Class (MDR) 5152100003L02 2305V004...
  • Page 16: Quattro

    Product description Quattro Electrical data 5452-51 5452100001 Nominal voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection switch, recommended setting Type of protection IP 24 Mains fuse* Max. permissible mains impedance in £ 0.24 accordance with EN 61000‑3‑11** Circuit breaker fuse characteristics B, C or D in accordance with EN 60898-1 Mains impedance at 6 switching cycles per hour.
  • Page 17 Product description Classification Medical Device Class (MDR) 5152100003L02 2305V004...
  • Page 18: Quattro Tandem

    Product description Quattro Tandem Electrical data 4682-52 4682-53 Nominal voltage 230 / 3~ 400 / 3~ Mains frequency Nominal current at 8 bar (0.8 MPa) 15.2 16.6 Motor protection switch, recommended setting Type of protection IP 24 IP 24 Mains fuse* Max.
  • Page 19: Filter

    Product description Filter Filter mesh size Air intake filter µm Fine filter µm Virus bacteria filter µm 0.01 Coalescence filter µm 0.01 Ambient conditions Ambient conditions during storage and transport Temperature °C -10 - +55 Relative humidity < 95 Ambient conditions during operation Temperature °C +10 - +40...
  • Page 20: Distance Between Rubber Feet

    Product description Distance between rubber feet Distances between the rubber feet for different pressure vessel volumes: Compressor unit type plate Membrane drying unit The type plate of the membrane drying unit is located on the side of the membrane drying unit. a (cm) b (cm) 20 l...
  • Page 21: Operation

    Product description Operation Duo with membrane drying unit Compressor unit Air intake filter Pressure vessel Condensate drain valve Non-return valve Pressure switch Safety valve Pressure gauge/display Pressure relief valve Cooler with ventilating fan Membrane drying unit Pressure limiting valve Humidity display Fine or virus bacteria filter Rinsing nozzle Membrane fibre...
  • Page 22: Assembly

    Assembly Assembly ≥ 40 °C Requirements The unit must not be set up or operated within the vicinity of the patients (within a radius of 1.5 m). The unit can be installed either at the same level as the surgery room or on a floor below (e.g. cel- lar).
  • Page 23: Transport

    Assembly Transport Quattro Tandem assembly and installation WARNING For weight reasons the unit is not supplied fully Risk of explosion of the pressure tank assembled. Instead, the compressor units are and pressure hoses installed at the site of use. The pressure tank and the pressure ❯...
  • Page 24: Installation

    Assembly Installation Remove the transport locks For safe transport, the unit is secured with foam padding blocks and a transport strap. Cut and remove the transport strap. ❯ Remove the foam padding blocks. ❯ Establishing the compressed air connection The supplied flexible pressure hose between the pipe system and the com- pressor prevents vibrations from being transmitted and thus reduces noise.
  • Page 25: Commissioning

    Assembly 10 Commissioning Make sure that none of the electrical cables ❯ leading to the unit are under any mechanical In many countries technical medical prod- tension. ucts and electrical devices are subject to Before initial start-up check that the mains sup- ❯...
  • Page 26: Checking The Switch-On/Cut-Off

    Assembly Measure the maximum current consumption Close the condensate drain valve as soon as ❯ ❯ (this is the value just before the cut-off pressure all of the accumulated condensation water has is reached). been blown out. If the reading deviates from the recommended setting then the motor protection switch needs to be adjusted (see "11.2 Adjusting the motor protection switch").
  • Page 27: Adjustment Options

    Assembly 11 Adjustment options Adjust the start-up pressure via the pressure ❯ difference ∆p at the adjustment screw. The pressure difference increases in the "+" 11.1 Adjusting the pressure switch arrow direction and decreases in the "-" arrow direction. WARNING The maximum permissible pressure difference Risk of explosion of the pressure ves- must not be set to more than 3 bar.
  • Page 28 Assembly Controller Remove the cover from the controller. ❯ Adjust the motor protection switch with the ❯ adjustment screw to the measured value (observe the range between the MIN permissi- ble setting and the MAX permissible setting of the motor protection switch, see "4 Technical data").
  • Page 29: Circuit Diagrams

    Assembly 12 Circuit diagrams 12.1 1/N/PE AC 110-127 V, 230 V layout 1 3 5 P> I > I > I > 2 4 6 L PE N Z2/U1 U2 Z1 Mains connection L/N/PE AC 230 V Pressure switch Compressor unit Fan motor, membrane drying unit Fan motor, noise insulation (if required) 12.2...
  • Page 30: 3/N/Pe Ac 230 V Layout, Quat

    Assembly 12.3 3/N/PE AC 230 V layout, Quattro Tandem 2 4 6 4 4 4 5 5 5 2 2 1 1 P> 1 3 5 2 4 6 2 4 6 14 22 14 22 1 3 5 13 21 13 21 2 4 6 I>...
  • Page 31: Usage

    Usage Usage 13 Operation Prior to working on the unit or in case of danger, disconnect it from the mains. 13.1 Switching the unit on/off Switch on the unit at the pressure switch by ❯ rotating it to the position "I AUTO". The compressor unit will start up automatically and fill the pressure tank.
  • Page 32: Maintenance

    Usage 14 Maintenance Prior to working on the unit or in case of danger, disconnect it from the mains. WARNING Risk of infection due to burst filters Particles enter the compressed air network and can therefore enter the mouth of the patient. Replace filters in accordance with the maintenance schedule.
  • Page 33: Changing The Filter

    Usage Fine or virus bacteria filter 14.3 Changing the filter Unscrew and remove the filter cover. ❯ Remove the filter. ❯ Prior to working on the unit or in case of Insert a new filter. danger, disconnect it from the mains. ❯...
  • Page 34: Checking The Safety Valve

    Usage 15 Taking out of use 14.4 Checking the safety valve The functioning of the safety valve must be 15.1 Taking the unit out of use checked at regular intervals in accordance with national regulations. If the unit is not to be used for a prolonged period of time, we recommend that it is properly WARNING shut down and taken out of use.
  • Page 35: Storage Of The Unit

    Usage 15.2 Storage of the unit WARNING Risk of explosion of the pressure tank and pressure hoses The pressure tank and the pressure ❯ hoses must be vented before they are stored or transported. Protect the unit against moisture, dirt and ❯...
  • Page 36: Troubleshooting

    Troubleshooting Troubleshooting 16 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. Prior to working on the unit or in case of danger, disconnect it from the mains. Error Possible cause Remedy...
  • Page 37 Troubleshooting Error Possible cause Remedy Knocking or loud noises on Compressor unit defective Disconnect all power from the ❯ device and inform a service the compressor technician. Reduced delivery. Compres- Air intake filter dirty Replace the air intake filter at ❯...
  • Page 38: Appendix

    Appendix Appendix 17 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)
  • Page 39 Appendix 5152100003L02 2305V004...
  • Page 40: Country Representatives

    Appendix 18 Country representatives Country Address UK Responsible Person: Duerr Dental (Products) UK Ltd. 14 Linnell Way Telford Way Industrial Estate Kettering, Northants NN 16 8PS Уповноважений представник в Україні: Приватне підприємство “Галіт” вул. 15 квітня, 6Є, с. Байківці, Тернопільський р-н, 47711, Україна...
  • Page 44 Hersteller / Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...

Table of Contents