18. Fault elimination

18.1 Flash code of the status display of the control panel

Flash code
1 x flash
2 x flash
3 x flash
4 x flash
5 x flash
Fault elimination
Meaning
Module defect.
Accessories fault.
Lifter raised.
Fault on the right motor.
Loose/defect connection
or motor defect.
Fault on the left motor.
Loose/defect connection
or motor defect.
Fault/braking fault on the
left motor.
Loose/defect connection
or motor defect.
Motors disengaged.
Immediate action
-
-
Completely lower lifter
Check connectors
Check connectors
Check connectors
Engage motors. Switch
control panel off and
then on again.
Further action
Speak to specialist dealer
Speak to specialist dealer
-
Speak to specialist dealer
Speak to specialist dealer
Speak to specialist dealer
-