Electromagnetic Compatibility (Emc); Information Regarding Electromagnetic Compatibility - Paravan PR Biolution User Manual

Table of Contents

Advertisement

Manufacturer's declaration - electromagnetic immunity
Biolution is designed to be operated under the electromagnetic conditions specified below. The customer or
the user of the device should ensure that it is used under these conditions.
Interference
immunity tests
Electrostatic
discharge (ESD) in
accordance with IEC
61000-4-2
Fast transient electric
bursts in accordance
with IEC 61000-4-4
Surges
in accordance with
IEC 6100-4-5
120
IEC 60601- test level
± 6 kV Contact
discharge
± 8 kV Air discharge
± 2 kV for power
cables
± 1 kV for input and
output cables
± 1 kV differential
mode voltage
± 2 kV common-mode
voltage
Compliance level
± 6 kV Contact
discharge
± 8 kV Air discharge
± 1 kV for input and
output cables
Not usable
Electromagnetic conditions -
guidelines
Floors should be made from wood
or concrete, or have ceramic
tiles. If the flooring is made from
synthetic material, the relative
humidity must be at least 30%.
The quality of the power supply
should correspond to that of
a typical business or hospital
environment.
The quality of the power supply
should correspond to that of
a typical business or hospital
environment.

Electromagnetic compatibility (EMC)