Paravan PR Biolution User Manual page 121

Table of Contents

Advertisement

Voltage dips, short
interruptions and
in the case of
fluctuations, the
power supply in
accordance with
IEC 61000-4-11
Magnetic field in
the power supply
frequency (50/60 Hz)
in accordance with
IEC 61000-4-8
NOTE UT is the mains voltage before using the test level.
Table 202 in accordance with DIN EN 60601-1-2, 6.8.3.201
Electromagnetic compatibility (EMC)
< 5% U
(>95% dip of
T
UT for 0.5 periods
40% U
(60% dip of
T
UT) for 5 periods
70% U
(30% dip of
T
U
) for 25 periods
T
< 5% U
(>95 % dip of
T
U
for 5 seconds
T
3 A/m
Not usable
3 A/m
The quality of the power supply
should correspond to that of
a typical business or hospital
environment.
If the Biolution user requires
continued functionality upon the
interruption of the power supply,
discharging from an uninterrupted
power supply or battery is
recommended.
Magnetic fields in the mains
frequency should correspond
to the typical values which are
present in a business and hospital
environment.
121