Elcometer 355 Operating Instructions Manual page 29

Coating thickness gauge
Hide thumbs Also See for 355:
Table of Contents

Advertisement

R
8.3 COUNTED AVERAGE MODE
The user predetermines how many readings should be taken to produce the average. This number is set
within the SEL mode of Setup.
As readings are taken, the reading number display shows 'n' and the number of readings taken so far. The
[
] symbol indicates that the main display is a true average. When the pre-determined count is reached,
a double beep will sound (the second beep will be extended and the red LED will show if the stored reading
is out of limits). At this point, the reading number displays the number of averages stored, and the main
display shows the last average saved. The statistics keys will show the batch statistics at all times in this
mode.
27

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents