6.2.7 F7 -A/D COUNT

To enter this parameter, press the [Tare] key when "
This parameter allows you to view the A/D counts from the internal A/D converter. This can
be an aid to service.
Press the [Tare] key to return to the PARAMETER menu.
Press the [Zero] key to return to weighing.
Typical value at zero is 30,000-90,000 (approx.)
Typical value at full capacity is 500,000 (approx.)
6.2.8 F8 –ZERO
This parameter allows the normal method of setting zero.
To enter this parameter, press the [Tare] key when "
The display will show the current settings, "
Press the [U.Wt./Units] key to change and press [Tare] to accept the value.
Press [Zero] to return to weighing.
6.2.9 F9 –PIN
This parameter allows setting of password to enter these Service Parameters.
To enter this parameter, press the [Tare] key when "
© Adam Equipment Company 2007