Enter Wiring Group Menu
From the main screen of Phasor select B (Wiring Group).
Select the Phasor Data to Display
1. In Wiring Group, use the left and right arrow ( O P ) keys to select a channel or wiring group in
Channel Configuration. The wiring selection of the selected channel/wiring group is highlighted.
As you move the cursor, you can observe that the data displayed in the Phasor screen is updated
instantly according to the current options.
2. Push Q (Back) on the control panel to go back to the main screen of Phasor.
The wiring group currently available for display depends on the wiring selection. See
Set the Wiring Selection

Select Circuit Type

1. From the main screen of Phasor select C (Y-Δ).
2. Push C (Y-Δ) to switch between the star (Y) and the triangle (Δ) connection. The current
connection type is displayed on the tab page.
The Y-Δ option is only valid when a 3-phase wiring group is selected. When Y is
selected, the phasor table displays the voltage of each phase, and the phasor diagram
is plotted in in the form of Y. When Δ is selected, the phasor table displays the voltage
of each line, and the phasor diagram is plotted in the form of Δ.

Zoom Phasor

In the Phasor screen, the Analyzer automatically determines the magnitude of each phasor in phasor
diagram based on the current measurements. However, the user can also manually zoom the phasor
diagram of the voltage and current for ease of viewing.
Enter the Zoom Menu:
From the main screen of Phasor select E (Zoom).
The options are as shown in
Note
for details.
Note
Table
19.
Power Analyzer
Operation
47