Why Do Maintenance; Keeping Records; Quality Assurance; Maintenance Task Schedule - GE LOGIQ C3 Premium Basic Service Manual

Hide thumbs Also See for LOGIQ C3 Premium:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5341787-100, R
IRECTION
Section 10-2

Why do Maintenance

10-2-1

Keeping Records

It is good business practice that ultrasound facilities maintain records of quality checks and corrective
maintenance. The Ultrasound Inspection Certificate (provided on page 10-16) provides the customer
with documentation that the ultrasound scanner is maintained on a periodic basis.
A copy of the Ultrasound Periodic Maintenance Inspection Certificate should be kept in the same room
or near the scanner.
10-2-2

Quality Assurance

In order to gain accreditation from organizations such as the American College of Radiology (USA), it
is the customer's responsibility to have a quality assurance program in place for each scanner. The
program must be directed by a medical physicists, the supervising radiologist/physician or appropriate
designee.
Routine quality control testing must occur regularly. The same tests are performed during each period
so that changes can be monitored over time and effective corrective action can be taken.
Testing results, corrective action and the effects of corrective action must be documented and
maintained on the site.
Your GE service representative can help you with establishing, performing and maintaining records for
a quality assurance program. Please contact us for coverage information and/or price for service.
Section 10-3

Maintenance Task Schedule

10-3-1
How often should care & maintenance tasks be performed?
The Care & Maintenance Task Schedule (provided on page 10-3) specifies how often your LOGIQ C3/
C5 Premium should be serviced and outlines items requiring special attention.
NOTE:
It is the customer's responsibility to ensure the LOGIQ C3/C5 Premium care & maintenance is
performed as scheduled in order to retain its high level of safety, dependability and
performance.
Your GE Service Representative has an in-depth knowledge of your LOGIQ C3/C5 Premium ultrasound
scanning system and can best provide competent, efficient service. Please contact us for coverage
information and/or price for service.
The service procedures and recommended intervals shown in the Care & Maintenance Task Schedule
assumes that you use your LOGIQ C3/C5 Premium for an average patient load (10-12 per day) and not
use it as a primary mobile unit which is transported between diagnostic facilities.
NOTE:
If conditions exist which exceed typical usage and patient load, then it is recommended to
increase the maintenance frequencies.
10 - 2
8
EVISION
Section 10-2 - Why do Maintenance
LOGIQ C3/C5 P
REMIUM
B
S
M
ASIC
ERVICE
ANUAL

Advertisement

Table of Contents
loading

Table of Contents