Transducer Connection - GE LOGIQ C3 Premium Basic Service Manual

Hide thumbs Also See for LOGIQ C3 Premium:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5273008-100, R
IRECTION
3-5-2-1
Back-end Processor Power Down (cont'd)
2.) The System-Exit window is displayed.
3.) Using the Trackball or Select key, select Shutdown.
4.) The shutdown process takes a few seconds and is complete when the power status LED is turned
amber.
5.) Disconnect the probes.Clean or disinfect all probes as necessary. Store them in their shipping
cases to avoid damage.
3-5-2-2
Scanner Shutdown
Disconnect the Mains Power Cable if necessary. For example: Relocating the scanner.
3-5-3

Transducer Connection

1.) Plug the probe connector into the probe port, then lock the probe.
NOTE:
Please ensure that the probe latch is in an unlocked position before you connect the probe to the
system.
NOTE:
It is not necessary to turn OFF power to connect or disconnect a probe.
8
EVISION
Figure 3-5 System Exit Window
Chapter 3 System Setup
VIVID FIVE
B
S
M
ASIC
ERVICE
ANUAL
3 - 9

Advertisement

Table of Contents
loading

Table of Contents