GE H
EALTHCARE
D
5341787-100, R
IRECTION
4-3-2-1Back-end Processor Power Down (cont'd)
2.) The System-Exit window is displayed.
3.) Using the Trackball or Select key, select Shutdown.
4.) The shutdown process takes a few seconds and the power off sequence is complete when the
power status LED is turned amber.
5.) Disconnect the probes.Clean or disinfect all probes as necessary. Store them in their shipping
cases to avoid damage.
4-3-2-2
Scanner Shutdown
Disconnect the Mains Power Cable is necessary. For example: Relocating the scanner.
DO NOT unplug and/or transport the unit until after the power off sequence has been completed.
CAUTION
Failure to do so may result in corrupted patient files.
8
EVISION
Figure 4-20 System Exit Window
Chapter 4 Functional Checks
LOGIQ C3/C5 P
REMIUM
B
S
M
ASIC
ERVICE
ANUAL
4 - 3