Safety Reminders

WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
DANGER
AC GROUND LINE (I.E., METER'S GROUND SWITCH IS OPEN), DON'T TOUCH
THE UNIT!
Two people should unpack the unit because of its weight. Two people are required whenever
CAUTION
a part weighing 19kg (35 lb.) or more must be lifted.
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate
CAUTION
to its operating environment.
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use
CAUTION
a three to two prong adapter. This defeats safety grounding.
Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is
CAUTION
present.
Do not use a 20 Amp to 15 Amp adapter on the 120 Vac unit's power cord. This unit requires
CAUTION
a dedicated 20 A circuit and can have a 15A plug if the on board peripherals do not cause the
unit to draw more than 14.0 amps.
Do not operate this unit unless all board covers and frame panels are securely in place.
CAUTION
System performance and cooling require this.
OPERATOR MANUAL(S)
CAUTION
The User Manual(s) should be fully read and understood before operating the LOGIQ C3/C5
Premium and kept near the unit for quick reference.
3 - 4
8
EVISION
Figure 3-1 Environmental Labels
Section 3-2 - Setup Reminders