Electrical Requirements
2-2-2-1
LOGIQ C3/C5 Premium Power Requirements
The following power line parameters should be monitored for one week before installation. We
recommend that you use an analyzer Dranetz Model 606-3 or Dranetz Model 626:
Table 2-3
Voltage Range
Power
Line Frequency
Power Transients
Decaying Oscillation
2-2-2-2
Inrush Current
Inrush Current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
2-2-2-3
Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be ready accessible.
POWER OUTAGE MAY OCCURE.
CAUTION
The LOGIQ C3/C5 Premium requires a dedicated single branch circuit. To avoid circuit overload
and possible loss of critical care equipment, make sure you DO NOT have any other equipment
operating on the same circuit.
2-2-2-4
Site Power Outlets
A desiccated AC power outlet must be within reach of the unit without extension cords. Other outlets
adequate for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and
national electrical codes.
2-2-2-5
Unit Power Plug
If the unit arrives without the power plug, or with the wrong plug, you must contact your GE dealer or
the installation engineer must supply what is locally required.
2-2-2-6
Power Stability Requirements
Voltage drop-out
Max 10 ms.
Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient, including
line frequency, synchronous, asynchronous, or aperiodic transients.
8
EVISION
(cont'd)
Electrical Specifications for LOGIQ C3/C5 Premium
PARAMETER
100-120V~
220-240V~
All applications
All applications
All applications
All applications
Section 2-2 - General Console Requirements
AREA
500VA
500VA
MAX. 750 VA
50/60Hz (±2Hz)
Less than 25% of nominal peak voltage
for less than 1 millisecond for any type of
transient, including line frequency,
synchronous, asynchronous, or
aperiodic transients.
Less than 15% of peak voltage for less
than 1 millisecond.
LOGIQ C3/C5 P
REMIUM
B
S
M
ASIC
ERVICE
ANUAL
LIMITS
2 - 3