Table 3 Measurement data
Element
Definition
LAST MEAS CHANNEL Last channel measured.
LAST MEAS TIME
Time of last measurement.
LAST ABS
Last absorbance reading.
LAST CONC
Concentration of last measurement.
NEXT MEAS TIME
Time the next measurement will be taken.
DARK
Number of A/D counts measured when the LED is
turned off.
REF
Reference A/D count used to compensate for natural
color and turbidity.
SAMPLE
A/D counts measurement (after color development)
used to determine the concentration of the sample.
DARK STD DEV
Standard deviation of dark counts out of 6 readings.
REF STD DEV
Standard deviation of reference counts out of
6 readings.
SAMPLE STD DEV
Standard deviation of sample counts out of 6 readings.
SAMPLE VOLUME
Total sample flush volume through the colorimeter for
the measurement cycle.
REAGENT 1
Calculated reagent delivery time to the sample based
on temperature, pressure and viscosity.
REAGENT 2
Calculated reagent delivery time to the sample based
on temperature, pressure and viscosity.
REAGENT 3
Calculated reagent delivery time to the sample based
on temperature, pressure and viscosity.
View instrument information
1. Select INSTRUMENT INFORMATION.
2. Select an option.
Option
ANALYZER INFO
MODULE INFO (option is
available when a module is
installed)
Set up LINK2SC
The LINK2SC procedure is a secure method for the data exchange
between process probes and LINK2SC-compatible photometers. Use an
SD memory card for the data exchange. Refer to the LINK2SC
documentation on
http://www.hach.com
LINK2SC procedure.
1. Select LINK2SC.
2. Select an option.
Option
Description
CREATE A NEW
Starts the grab sample operation for the measurement
JOB
value exchange between the analyzer and the
laboratory.
JOB LIST
Selects the job file to send the job to the laboratory or
deletes the job. JOB TO LAB—the analyzer data is sent
to the SD card as a job file. ERASE JOB—deletes the
data.
JOB ID MIN
Specifies the minimum value for the ID number range.
JOB ID MAX
Specifies the maximum value for the ID number range.
Use an SD card
Use an SD memory card to update the software and firmware and to
download the event and data logs. The SD icon is visible in the upper
status bar of the main measurement screen when a card is installed. The
Description
Shows the software information and
serial number.
Shows the connected modules with
software information and serial number.
for a detailed description of the