Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification
Details
Dimensions (W x D x H)
452 x 360 x 804 mm (17.8 x 14.2 x 31.7 in.)
Enclosure
Rating: NEMA 4x/IP65
Material: PC/ABS case, PC door, PC hinges and
latches, 316 SST hardware
Indoor use only. Keep away from direct sunlight.
Weight
20 kg (45 lb) without reagents and standards,
36.3 kg (80 lb) with reagents
Mounting
Wall, panel or table
Protection class
I
Pollution
2/II
degree/installation
category
Power requirements
AC: 100–240 VAC, 50/60 Hz
Instrument: 0.5 A nominal, 2.6 A maximum
Connection: 18 to 16 AWG wire, 18 AWG stranded
recommended
Operating temperature
5 to 45 °C (41 to 113 °F)
Operating humidity
5 to 95% non-condensing
–20 to 60 °C (–4 to 140 °F)
Storage temperature
Table 1 General specifications (continued)
Specification
Details
Maximum altitude
2000 m (6560 ft)
Four; load impedance: 600 Ω maximum
4–20 mA outputs
Connection: 22 to 16 AWG wire, 22 to 20 AWG
recommended, twisted pair shielded wire
Alarm relay outputs
Four; type: not powered SPDT relays, each rated
at 5 A resistive, 240 VAC maximum
Connection: 18 to 16 AWG wire, 18 AWG stranded
recommended
Digital inputs
Four; connection: 22 to 16 AWG wire, 22 to
20 AWG stranded (isolated DC voltage input or an
open-collector/relay contact closure input)
recommended
Fuses
Input power—AC: T 1.6 A, 250 VAC; DC: T 6.3 A,
250 VAC
Output power—AC: T 5.0 A, 250 VAC; DC: T 1.6 A,
250 VAC
Alarm relay outputs: T 5.0 A, 250 V
Fittings
Sample line and sample bypass drain: 6 mm OD
push-to-connect fitting for plastic tubing
Air purge air inlet: 6 mm OD push-to-connect fitting
for plastic tubing
Chemical and case drains: 11 mm (7/16 in.) ID slip-
on fitting for soft plastic tubing
Pressure: 2–87 psi to preset pressure regulator
Sample pressure, flow rate,
and temperature
Flow rate: 55–300 mL/minute
Temperature: 5 to 50 °C (41 to 122 °F)
Number of sample streams
1, 2, 4 or 6; programmable sequence
Air purge (optional)
0.425 m
Certifications
ETL certified to UL and CSA standards, CE marked
1
standard cubic feet per hour
3
1
/hour (15 scfh
), instrument quality air
English 3