Hach Polymetron 9610sc SiO2 Installation Manual page 34

Hide thumbs Also See for Polymetron 9610sc SiO2:

Advertisement

Option
Description
OFF DELAY
Sets a delay time (0–300 seconds) to set the relay off
(default = 5 seconds).
ON DELAY
Sets a delay time (0–300 seconds) to set the relay on
(default = 5 seconds).
4. Select FEEDER CONTROL in the FUNCTION menu and then select
the applicable options in the ACTIVATION menu.
Option
Description
PHASE
Specifies the relay status if the process value is larger
than the setpoint. HIGH(default)—sets the relay to on
when the process value is larger than the setpoint. LOW
—sets the relay to on when the process value falls below
the setpoint.
SET SETPOINT
Sets the process value for the relay to toggle between the
high and low value (default = 10).
DEADBAND
Sets a delay so that the relay will be stable as the
process value converges to the setpoint.
OVERFEED
Sets the maximum amount of time to get to the process
TIMER
setpoint. When the time expires and the relay does not
show the setpoint, the relay is set to off. After an overfeed
alarm occurs, reset the timer manually.
OFF DELAY
Sets a delay time to set the relay to off (default =
5 seconds).
ON DELAY
Sets a delay time to set the relay to on (default =
5 seconds).
5. Select EVENT CONTROL in the FUNCTION menu and then select
the applicable options in the ACTIVATION menu.
Option
Description
SET SETPOINT Sets the value to set the relay to on.
DEADBAND
Sets a delay so that the relay will be stable as the process
value converges to the setpoint.
OnMax TIMER
Sets the maximum time the relay stays on (default =
0 min).
Option
Description
OffMax TIMER
Sets the maximum time the relay stays off (default =
0 min).
OnMin TIMER
Sets the time the relay stays on, independent from the
measured value (default = 0 min).
OffMin TIMER
Sets the time the relay stays off, independent from the
measured value (default = 0 min).
6. Select SCHEDULER in the FUNCTION menu and then select the
applicable options in the ACTIVATION menu.
Option
Description
HOLD OUTPUTS Holds or transfers outputs for the selected channels.
RUN DAYS
Selects the days the relay stays on. Options: Sunday,
Monday, Tuesday, Wednesday, Thursday, Friday,
Saturday
START TIME
Sets the start time.
INTERVAL
Sets the time between activation cycles (default = 5 min).
DURATION
Sets the period of time the relay is set to on (default =
30 sec).
OFF DELAY
Sets the time for additional hold/output time after the
relay is set to off.
7. Select WARNING in the FUNCTION menu and then select the
applicable options in the ACTIVATION menu.
Option
Description
WARNING LEVEL Sets the level for the warning activation and starts
applicable individual warnings.
English 15