• GRAB SAMPLE OUT: This option is used to extract a sample directly
from a sample line for external analysis.
Measure a grab sample or standard
Use the grab sample funnel to measure grab samples taken from other
areas in the system or to measure a standard for a calibration
verification.
1. Collect 250–500 mL of the sample or standard in a clean container.
2. Rinse the funnel outside of the unit with the collected sample.
3. Install the funnel again.
4. Select GRAB SAMPLE>GRAB SAMPLE IN.
5. Complete the instructions on the display.
Get a grab sample from the analyzer
Use the grab sample tube to manually dispense a grab sample from one
of the sample sources for external analysis.
The sample source is measured by the analyzer immediately after the
sample is dispensed. The value of the sample source and the job
identification number for the grab sample are shown on the display.
1. Select GRAB SAMPLE>GRAB SAMPLE OUT.
2. Complete the instructions on the display.
3. Put the grab sample out tube in a clean container. The grab sample
tube is located in the lower unit of the cabinet on the left side.
4. Push down on the spigot (upper left corner of the lower
compartment) for the grab sample tube to dispense a grab sample.
Set up the system
The configuration settings can be changed in the SETUP SYSTEM
menu or in the analyzer setup. Refer to
on page 9.
1. Select SETUP SYSTEM.
Start the analyzer setup
2. Select one of the options.
Option
Description
MEAS MODE
Changes the mode of the measurement cycle.
Options: interval or continuous (default). The
continuous mode measures approximately every
9 minutes.
EDIT INTERVAL
Changes the interval time when MEAS MODE is set
(conditional)
to interval. Options: 10–240 minutes (default =
15 minutes).
MEAS UNITS
Changes the measurement units that are shown on
the display and in the data log. Options: ppb
(default), ppm, mg/L, µg/L.
SIGNAL AVERAGE
Selects the number of measurements used to
calculate an average measurement (1–5). This
reduces variability in measurements (default = 1) (no
average).
EDIT ANALYZER
Changes the name that is shown on the top of the
NAME
measurement screen (maximum of 16 characters).
EDIT CHANNEL
Changes the name of the sample source that is
NAME
shown on the measurement screen (maximum of
10 characters).
CONFIGURE
Starts or stops measurements for individual sample
SEQUENCER
sources. Sets the measurement order of the sample
(optional)
sources when there is more than one sample
source.
SET DATE & TIME
Sets the analyzer time and date.
DISPLAY SETUP
Changes the language. Adjusts the order
measurements are shown. Adjusts the display
contrast settings.
DISABLE
Stops scheduled maintenance alerts for individual
REMINDERS
components. Options: tubing, stir bar, colorimeter
cell, air filter, stir motor, air relief valve, pinch valve,
air compressor, reagent valves, sample valves,
standard valves, colorimeter LED, fan filter, air
check valve.