AMC‑15 Mobile Dental Cart OPERATION & MAINTENANCE MANUAL...
Page 2
• CAUTION: Damage to property or the environment may result if ignored. • NOTE: Important additional information and hints. Aseptico accepts no liability for direct or consequential injury or dam- s CAUTION: Do not run saline solutions through the water sys- age resulting from improper use, arising in particular through the tem—saline will corrode the plumbing.
AMC‑15 SETTING UP THE UNIT Inside the waste compart- ment, make sure that the Unpack the AMC-15 from its cable from the waste tank cardboard crate following sensor is plugged into the the PACKING INSTRUCTIONS connector labeled WASTE on page 16, in the reverse SENSOR on the back wall order.
AMC‑15 SYSTEM OPERATION Connect the power cord to the power inlet on the rear of the cart, then plug the other end of the power cord into a FOOT CONTROL — Press grounded electrical outlet of the proper voltage correspond- the foot pedal to acti- ing to the unit version (115V or 230V).
Page 10
AMC‑15 SYSTEM OPERATION (Continued) AIR COOLANT — Air coolant (chip air) works in combina- WATER BOTTLES — The AMC-15 incorporates a self-con- tion with the water coolant tained pressurized water system which includes: two 1-liter setting to create a fine mist water bottles, a bottle select switch and a bottle pressure for the working handpiece.
Replacement amalgam sep- clamp , then direct the arators (PN 730595) are avail- waste hose to a suitable able through Aseptico. Refer drain. to the label on the amalgam c. Switch ON the waste purge pump to empty the waste separator for the address of tank.
Page 12
Toggle ON the bottle pressure switch. 80 psi (5.5 bar). Pressure is controlled by the system pressure regulator . Contact Aseptico customer service in the unlikely f. Purge all water lines simultaneously or event that adjustment in necessary. one at a time: Hold handpieces over a...
REPLACEMENT PARTS LIST are tight. • Check that water bottle pressure toggle is in the ON position. The following items can be purchased from Aseptico for replace- Insufficient vacuum: ment (or as spares) of standard and optional equipment. • Check vacuum hose assemblies for blockage.
AMC‑15 SYMBOL DEFINITIONS Sterilizable up to the temperature indicated Waste Tank full -or- Waste Level Sensor Cable Caution – Consult Accompanying Documents disconnected Consult Instructions For Use SPECIFICATIONS Do Not Use If Damaged Cart Size 23 5” W x 30 0” L x 36 5” H (56 7 cm x 76 2 cm x 92 7 cm) Use By Date Cart Weight (with all optional equipment) 139 lb (63 kg)
(at its sole option and discretion) to repair or replace any Pneumatic Handpieces with integrated air and water cooling defective component or product in part or whole. Aseptico shall be 3-way air/water syringe the sole arbiter of such action.
AMC‑15 PACKING INSTRUCTIONS Insert the cardboard end panels and side panels into Remove all liquid wastes, water from water bottles, and the the crate base, making sure bypass filter and optional amalgam separator. the “End Up” labels pointing up. Use the adhesive Remove all accessories from the unit and purge air pressure strips on the inside of the end from air tank.