PERCENTAGE WEIGHING
AGS is equipped with percentage weighing function, please follow the below procedure to enable
the function and sampling method.
a. AGS is on and both Zero indicator and Stable indicator are on
b. Press MODE until display shows PCT as weighing unit on the display
c. Place the sample load that represents 100% of the value for percentage weighing on the
platter
d. Press and hold MODE, and the display shows "S=100"
e. The scale is now ready for percentage weighing. Put an object on the platter and the display
will record a percentage of the original sample.
CHECK WEIGHING OPERATION
AGS is equipped with check weighing function. Users can use this function to set high and low
limits to quickly check if the weight, PCS, or percentage is within the defined range.
SETTING HIGH AND LOW LIMITS
a. AGS is on and both Zero indicator and Stable indicator are on.
b. Use MODE to select appropriate weighing unit : metric units(g/kg), imperial unit (lb), PCS and
PCT (%)
c. Press and hold MR until display shows "CHKECY on" or "CHKECY off" , then continue holding MR
until it displays "Hi".
d. Enter value for low limit
Use M+ for value increase
Use TARE for value decrease
Use MR for shifting digits
Press MODE to confirm and proceed to low limit setting.
e. Display shows "LoW".
f. Enter value for low limit
Use M+ for value increase
Use TARE for value decrease
Use MR for shifting digits
Press MODE to save and exit.
ENABLING/DISABLING CHECK WEIGHING OPERATION
a. Check Weighing operation is engaged after completing high and low value setting.
b. To disable check weighing operation, press and hold MR until AGS displays "CHKECY off"
Need help?