Operation Manual – "R" Units

Tank Level Sensors

In addition to the external level indicators, the units are also equipped with electronic
level sensors. In case the Test Oil level is low or the Cleaning (MACC) fluid level is High,
the unit will STOP operation and popup a screen message.
Drain the Cleaning Fluid through the rear drain valve, or Add more Test Oil. You can
monitor the level sensors in the Service Screen (Settings).

Degassing the system

When Changing a Filter or Fluid, it is necessary to perform a Hydraulic System Degas:
Control Panel>>HOME>>Options>>Settings>>Machine>>HP Degas
This function will ensure that the hydraulic system is free of air.
This function should be repeated as many times necessary, if the user feels that the
system is not air free. If the system is still with air after 5 times, contact your service
provider.