OTB Start Up Procedure

Press SYSTEM DETAILS
Press UV LAMP TEST
UV-Lamp Intensity should be greater than
70%, if less replace the UV-Lamp.
Wipe out the inside of the suction cup with a
lint free cloth.
Empty the trap at the bottom of the wash
bowl. Use teflon tape on the trap plug and
retighten. Do not overtighten, it only needs
to be hand tight.
Verify the water quality sensor is illuminated
green, indicating the water quality is 1MΩ or
better.