TROUBLESHOOTING PROCEDURES

PROCEdURE 1.
Complete this procedure if the console cannot connect to ScanPoint
1. In ScanPoint, retry the connection by clicking the Find New button. Repeat this step up to 3 times.
If the console does not connect, continue to the next step.
2. Turn the console off, turn the console on, and then press the ScanPoint button
ScanPoint, click Find New.
If the console does not connect, continue to the next step.
3. On the PC, click Find New. While the device is attempting to connect to ScanPoint, remove the battery.
4. Reinsert the battery, allow the device to power on, and then press the ScanPoint button
5. On the PC, click Find New.
If the console does not connect, contact Verathon
Operations & Maintenance Manual: Troubleshooting
TROUBLESHOOT SCANPOINT CONNECTION
.
®
Customer Care.
®
. On the PC, in
.
65