11. If calibration is successful, a "Calibration Successful" message is displayed on the computer.
If calibration fails, a Calibration Failure message appears. Ensure that the calibration chamber has
sufficient water and that the probe is seated properly in the calibration lid, and then on the Calibration
Failure message, click Yes. ScanPoint
12. On the console, click the Exit button
communication with ScanPoint QuickPrint.
13. Remove the probe from the calibration lid, and then dry it with a clean, soft cloth.

DEVICE DISPOSAL

The AortaScan AMI 9700 and related devices may contain mineral oils, batteries, and other environmentally
hazardous materials. When the AortaScan AMI 9700 has reached the end of its useful service life, return
the device, battery charger/wireless hub, and related accessories to a Verathon
disposal. Alternatively, follow your local protocols for hazardous waste disposal.
Operations & Maintenance Manual: Cleaning & Maintenance
restarts the calibration.
®
. This terminates the calibration procedure and ends
Service Center for proper
®