Instruction Manual
7
Qualification Testing
7.1 Standards
The sensor NS-xx/DPG2-xxx complies with:
•
IEC/EN 60068-2-2
•
IEC/EN 60068-2-78
•
IEC/EN 60068-2-27
•
IEC/EN 60068-2-6
•
IEC/EN 60068-2-14
•
IEC/EN 61000-6-2
•
IEC/EN 61000-6-4
7.2 Periodical Testing
The periodical testing is done every 3 years in terms of a product audit.
7.3 Material Testing
All materials used in the process will be released by checking the corresponding supplier certificates
when available. A regular material analysis from an independent laboratory will not be scheduled.
MEAS Deutschland GmbH
Hauert 13, D-44227 Dortmund, Germany
DPG-Series Inclinometer
High Temperature Storage
Ta= 85°C, medium : air, 1000h
High Humidity
Ta= 85°C , RH= 85%, 1000h
Mechanical shock test
A=30g, t=11ms, halfsine, 50 cycles per axis
Vibration loading
10 to 150 Hz, 2.5mm amplitude, 5g const. acceleration,
1 octave/minute, 20 cycles per axis
Thermo shock
T
=85°C ,T
=-40°C,medium:air-air,T
High
Low
T
: 30sec,100 cycles
change
Interference resistance industry
Emitted interference industry
www.meas-spec.com