Qualification Testing; Standards; Periodical Testing; Material Testing - Measurement Computing DPG Series Instruction Manual

Inclinometer
Table of Contents

Advertisement

Instruction Manual
7

Qualification Testing

7.1 Standards

The sensor NS-xx/DPG2-xxx complies with:
IEC/EN 60068-2-2
IEC/EN 60068-2-78
IEC/EN 60068-2-27
IEC/EN 60068-2-6
IEC/EN 60068-2-14
IEC/EN 61000-6-2
IEC/EN 61000-6-4

7.2 Periodical Testing

The periodical testing is done every 3 years in terms of a product audit.

7.3 Material Testing

All materials used in the process will be released by checking the corresponding supplier certificates
when available. A regular material analysis from an independent laboratory will not be scheduled.
MEAS Deutschland GmbH
Hauert 13, D-44227 Dortmund, Germany
DPG-Series Inclinometer
High Temperature Storage
Ta= 85°C, medium : air, 1000h
High Humidity
Ta= 85°C , RH= 85%, 1000h
Mechanical shock test
A=30g, t=11ms, halfsine, 50 cycles per axis
Vibration loading
10 to 150 Hz, 2.5mm amplitude, 5g const. acceleration,
1 octave/minute, 20 cycles per axis
Thermo shock
T
=85°C ,T
=-40°C,medium:air-air,T
High
Low
T
: 30sec,100 cycles
change
Interference resistance industry
Emitted interference industry
www.meas-spec.com
14 of 15
July 2015
: 15min,
dwell
phone: +49-(0)231-9740-0
fax: +49-(0)231-9740-20

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DPG Series and is the answer not in the manual?

Questions and answers

Table of Contents