Wall-mounted version
5.2.2

Docking to the wall mount

1. Connect the potential equalization cable (1) to the potential equalization pin of
the wall connector (2) and the wall mount (3).
CAUTION
Risk of patient injury due to gas supply failure
The patient may be put at risk if the central gas supply system fails. In this case, the
oxygen supply can be ensured by connecting an alternative source of oxygen:
► By using the pressure reducer that is already installed.
► Alternatively, it is possible to use an additional compressed gas hose between
Atlan and the O
starting the device. The compressed gas hose must remain freely accessible at
all times. The compressed gas cylinder must be equipped with a pressure
reducer and the appropriate connector when opting for this setup.
► If neither of these options is feasible, the compressed gas hose for oxygen can
be detached from the central gas supply system and connected to a suitable
compressed gas cylinder.
38
Supplement
supply. Connect the hose to the rear of the device before