Electromagnetic conformity
HeartSine Gateway is suitable for use in all professional and domestic establishments. It is not
intended for use near intentional transmitters of radio energy such as high frequency surgical
equipment, radar installations or radio transmitters, nor in the vicinity of magnetic resonance
imaging (MRI) equipment.
HeartSine Gateway is intended for use in the electromagnetic environments specified in
the tables below. The user of HeartSine Gateway should assure that it is used in such an
environment.
The essential performance of HeartSine Gateway is the ability to periodically request and
receive data via the existing USB interface of the host defibrillator, and to wirelessly transmit
this data over an established Wi-Fi connection. Operation outside of the environments specified
below may result in improper operation of HeartSine Gateway device.
There are no special maintenance procedures required to ensure that the essential performance
and basic safety of HeartSine Gateway are maintained with regard to electromagnetic
disturbances over the service life of the device.
Emissions Test
RF CISPR 11
Harmonic Emission
IEC/EN 61000-3-2
Voltage Fluctuations/
Flicker Emission
IEC/EN 61000-3-3
50
Compliance
Electromagnetic Environment – Guidance
HeartSine Gateway uses RF energy only for its internal
Group 1 Class B
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
Not Applicable
equipment.
HeartSine Gateway is suitable for use in all establishments,
including domestic and those directly connected to the
Not Applicable
public low-voltage power supply network that supplies
buildings used for domestic purposes.