Chapter 5 Start and monitor a run
Monitor a run from the instrument
(Optional) View
the injection list,
change injection
settings or order,
or specify
replicates and re-
injections
42
In the Plate properties screen:
1. Touch Injection options.
2. Touch an injection group, then configure the injection list:
• Touch and drag an injection group to a new location in the injection list.
• Touch Inject first—Moves the selected injection group to the top of the
injection list.
• Touch Edit and re-inject—Adds replicates or re-injections to the injection list.
You can also modify Run module, Injection time, Injection voltage, Run
time, or Run voltage for these injections.
Note: Replicate indicates a duplicate—the replicate has the same parameters
as the original injection. Re-injection indicates a copy of the injection, but with
user-modified parameters.
Note: The changes to the run conditions apply only to the replicate or re-
injection. The changes are not saved to the run module.
3. Touch Done.
SeqStudio
Genetic Analyzer Instrument and Software Getting Started Guide