Load The Cbc, The Sample Plate, And The Cartridge - Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer Getting Started Manual

Create and run plate setups, monitor runs, and view results
Hide thumbs Also See for Applied Biosystems SeqStudio Genetic Analyzer:
Table of Contents

Advertisement

2
Chapter 2 Prepare the samples and the instrument
Prepare the instrument
Display
Yellow
OK for use, but:
• 1–25 injections remain for the capillary
• It will exceed the manufacturer's expiry
• It will exceed the maximum number of
Red
The cartridge has expired:
• No injections remain for the capillary
• It has exceeded the manufacturer's expiry
• It has exceeded the maximum number of
[1]
The maximum number of days allowed on the instrument is 120 days for the SeqStudio
180 days for the SeqStudio
Instrument and Software User Guide .
Load the CBC, the
sample plate, and
the cartridge
12
Cartridge
or
array,
or
date within 2 weeks,
days allowed on the instrument within
[1]
2 weeks
or
array,
or
date,
days allowed on the instrument
Genetic Analyzer Cartridge v2 (Cat. No. A41331). For more information, see
See SeqStudio
Genetic Analyzer Instrument and Software User Guide if a new
Cathode Buffer Container is required.
See SeqStudio
Genetic Analyzer Instrument and Software User Guide if a new
cartridge is required.
3. Touch Close, then touch
The Eject plate command is disabled for a few minutes after you insert a cartridge. If
you are loading the CBC, sample plate, and cartridge at the same time, you can save
time by loading the CBC and sample plate before you insert the cartridge.
For information on loading the individual components, see:
• "Prepare and load the plate or tubes" on page 14
• SeqStudio
Genetic Analyzer Instrument and Software User Guide
In the home screen:
SeqStudio
OK for use, but:
• 1–25 injections remain for the cathode
buffer,
• It will exceed the manufacturer's expiry
date within 2 days
• It will exceed the maximum number of
days allowed on the instrument (14 days)
within 2 days
The buffer has expired:
• No injections remain for the cathode
buffer,
• It has exceeded the manufacturer's expiry
or
date
• It has been installed on the instrument for
[1]
>14 days
Genetic Analyzer Cartridge (Cat. No. A33671) and
.
Genetic Analyzer Instrument and Software Getting Started Guide
Cathode buffer
or
or
or
SeqStudio
Genetic Analyzer