3.9 Setting Main Screen Time, Date, Temperature, and Concentration Units

3.10 Setting Maximum Range

3.11 System Calibration

16116Operation.fm
Table 5 Main Menu Option Descriptions*
MODEL
SITE
LOG
COMM
42.
Time, date, and units for temperature and concentration may be changed on
the Model 8450.
From the Main menu, place the magnet over the
and permits movement of the cursor. Move the cursor next to the date and
select
. The parameter will begin to blink. Use the arrow icons to scroll
E
through to the desired day then select
to select the desired month then year. When finished, select the
the same procedure to change the time (hours/minutes), temperature units
(°F or °C), and concentration units (ppm or mg/L).
Note: Parts per million (ppm) and milligram per liter (mg/L) are equivalent units, that is,
1.0 ppm equals 1.0 mg/L.
Prior to Span calibration, set the maximum range value to the highest chlorine
concentration that will be measured by the monitor. Setting the "RNG"
parameter (found in the "RDG" menu) helps optimize gain control during span
calibration.
Settings are: 5,10,15, and 20 (default). Greatest accuracy is obtained when
the Maximum Range setting is not set excessively higher than it can be
reasonably expected to occur at a specific site. For example, there is nothing
to be gained by setting the Maximum Range to 20 at a site which has never
exceeded 5 ppm and is not expected to do so in the future.
Calibration of the instrument is accomplished with two major procedures
known as ZERO and SPAN. If problems are experienced with calibration after
following these guidelines and procedures, see
Setting Main Screen Time, Date, Temperature, and Concentration Units
8450 Ver. #.##, ## ppm- Free Cl (display only)
Change site (site information)
Change site (16 locations)
SETTINGS
DATA
REPORT
Addr (###); not used
Baud (1200, 2400, 4800, 9600)
Par (Odd, even, none)
. Proceed using this same technique
E
Section 3
Rate (# min.)
Ctrl (off / on / Rst)
day mo.yr.hr:min
IND/HIGH/LOW/AVE
##ppm/ mg/L
DEVICE (term/print)
TYPE (none/long/short)
arrow icon. This activates
Up
icon. Use
E
section 4.3 on page
54.