Factory Default Parameters; Adjusting Display Screen's Contrast - Hach 8450 Instrument Manual

Residual free chlorine monitor
Table of Contents

Advertisement

Section 3

3.5 Factory Default Parameters

3.6 Adjusting Display Screen's Contrast

Page 30
Factory Default Parameters
Version
V1.3
Date
15 Jan 96
Time
08:00
Site
Sigma, MEDINA, NY
Range
20
Conc Units
ppm
Precision
0.01
Temp Units
°F
Damping
2 seconds
Alarm 1 Sp
2.00 ppm
Alarm 1 Rsp
1.00 ppm
Alarm 1 Delay
10 seconds
Alarm 1 Lfit
0000
Alarm 2 Sp
5.00 ppm
Alarm 2 Rsp
2.50 ppm
Alarm 2 Delay
10 seconds
Alarm 2 Lfit
0000
Check Delay
5 seconds
Check Lfit
0000
Vma Ch 1
Units, ppm
Vma Ch 1 Spn
20.0
Vma Ch 1 Volts
0–1 V dc
Vma Ch 1 Ma
4-20 mA
Vma Ch 2 Units
°F
Vma Ch 2 Spn
212
Vma Ch 2 Volts
0–1 V dc
Vma Ch 2 Ma
4-20mA
Gpot
478
Zero Conc
0.00
Span Conc
20.0
%rel.sens
100%
Password
000
Security
Disabled
Comm Address
1
Comm Baud
9600
Comm Parity
None
Dlog Rate
1 Minute
Dlog Control
Off
This adjustment requires two magnets and may be performed while viewing
any screen.
1. Hold a magnet on the
icon.
F
16116Operation.fm

Advertisement

Table of Contents
loading

Table of Contents