Connections And Setup - Hach 8450 Instrument Manual

Residual free chlorine monitor
Table of Contents

Advertisement

2.7.1

Connections and Setup

Figure 12
Top View of Pins and Terminal Block
Figure 13
Plug-in Connector in Place
Figure 14
Extension Leads and Milliampmeter Connections
16116Install.fm
Small straight-slot screw driver
Milliampmeter
Color-coded connection wire, 10 in. (25.4 cm)/test leads
4-20 mA Surge Protector(s) (Cat. No. 16326)
Warning: Remove ac power to the instrument before proceeding.
1. Remove electrical power to the meter.
2. Unscrew the four cover face plate screws and remove the cover.
3. Unscrew the upper and lower right screws that hold the display circuit
board to the meter and swing the hinged display board back to access the
main circuit board inside the case.
4. As shown in
Figure
12, insert the two pins of the Surge Protectors into
pins 7 and 8 (channel 1) and 5 and 6 (channel 2) on the terminal block.
5. Tighten the clamp screws on this connector.
6. Connect the wires from the color-coded wire into surge protector terminal
screws. Route the wires through the bottom of the case (under the
display) and label the wires for channel 1 and 2 and + and –.
7. Temporarily tape the ends of channel 2 wires to avoid the possibility of
shorts. Connect a milliampmeter to the channel 1 extension leads.
8. Plug the screw-terminal block with attached surge protectors installed into
TB4 connector on the 8400 meter's main circuit board
Note: The plug-in screw terminal connector is keyed to ensure that it plugs in correctly.
Note: The location of TB4 and its numbered terminals are marked on the main circuit
board.
9. Swing the display back into the closed position, using care to avoid
pinching or snagging wires
10. Power the meter on again and proceed to
(Figure
14).
section
2.7.2.
4-20 mA Surge Protector
Section 2
(Figure
13.)
Page 19