Connecting The Uv Sensor; Attaching The Wiper Motor; Installing The Terminal Box; Installing The Lamp Protection Tube - ProMinent Dulcodes M Operating Instructions Manual

Uv-system
Table of Contents

Advertisement

Assembly aid
Auxiliary
assembly tube
Protection tube
®
ProMinent
5.2.5

Connecting the UV Sensor

Guide the O-ring over the sensor thread as far as the seal groove.
Screw the sensor into the sensor connection and tighten gently.
Connect the UV sensor using the black sensor connection cable.
5.2.6

Attaching the Wiper Motor

Loosen and remove the nuts on the four inserted screws. Make sure that the gears do not
become detached from the motor.
Push the wiper motor carefully onto the quick coupler and secure it using the four screws.
NOTE
Do not cover the bayonet coupler with the safety shield until the lamp and its protection
tube have been installed.
5.2.7

Installing the Terminal Box

A terminal box containing the ignition device must be installed on the radiation chamber of
systems with a conventional ballast.
Unscrew the terminal box and remove the lid.
Secure the terminal box to the mounting panel on the support stirrup of the temperature
gauge using the screws provided.
Place the lid on the terminal box and screw it down.
5.3

Installing the Lamp Protection Tube

IMPORTANT
Examine the protection tube for damage prior to installation. Damaged tubes must not be
installed.
NOTE
The protection tube is inserted into the radiation chamber from the lid end as standard. If
there is more space on the motor side, making it easier to insert it from this side, a special
assembly aid is required (optional), see also 5.1.2 "Preparation for Assembly".
Push the protective tube carefully over the auxiliary assembly tube from the lid end until it
locks into place. The protective tube should now protrude by approximately 1 cm.
Push the O-ring provided over the lid end of the protection tube.
Check whether the Teflon ring is in the correct groove on the clamp. If not, push the Teflon
ring into the clamp until it jumps into the correct groove.
Push the O-ring provided over the protection tube.
Push the clamp over the protection tube and screw it into the radiation chamber; tighten it
by hand.
Remove the auxiliary assembly tube and unscrew the assembly aid.
Push the O-ring provided over the motor side of the protection tube.
Push the clamp over the protection tube on the motor side and screw it into the radiation
chamber; tighten it by hand.
Use the hook spanner to tighten the tube clamp on both sides. Only a small amount of force
is required.