Commissioning
Page 30
Use the UP or DOWN button to set to ON and confirm by pressing ENTER. The message
appears "Data has been saved".
Press the CHANGE button to go to the "MANUAL or EXTERNAL CONTROL" display.
Confirm by pressing ENTER, the setting to be adjusted now flashes.
Use the UP or DOWN button to set to MANUAL and confirm by pressing ENTER. The
message appears "Data has been saved".
Press the CHANGE button to go to the MANUAL CONTROL display.
Confirm by pressing ENTER, the reference value to be adjusted now flashes.
Use the UP or DOWN button to set the required reference value and confirm by pressing
ENTER. The message appears "Data has been saved".
Press the CHANGE button to go to the SAFETY THRESHOLD display.
Confirm by pressing ENTER, the safety threshold value to be adjusted now flashes.
Use the UP or DOWN button to set the required safety threshold and confirm by pressing
ENTER. The message appears "Data has been saved".
Press the CHANGE button to go to the WARNING THRESHOLD display.
Confirm by pressing ENTER, the warning threshold value to be adjusted now flashes.
Use the UP or DOWN button to set the required warning threshold and confirm by pressing
ENTER. The message appears "Data has been saved".
Use the START/STOP button to re-start the system.
E.g.:
UV treatment is based on operating data for the end of the lifetime of the radiator, which is
indicated by a 50 % radiation output. In order to achieve a constant radiation output greater
than 50 %, the reference value must be set at 60 %, the safety threshold at –10 [%] and the
warning threshold at –5 [%]. Regardless of the lifetime of the radiator, the radiation output is
now automatically readjusted so as to constantly obtain a probe signal equal to 60 % of the ori-
ginal value. If the signal falls short of this value by 5 per cent, a warning message will be issued.
With a 10 per cent shortfall an alarm message is issued and the system shuts down, as now the
minimum radiation output has fallen to 50 %, which was the level defined when designing the
system.
6.3.3.2 Setting the radiation output in operating mode "external control"
Before commissioning the operating mode "external control ", a definitive reference value zzz
needs to be calculated for the system:
zzz = reference value x F/F
Press the CHANGE button to go to the CONTROL display.
Confirm by pressing ENTER, REQUEST RELEASE CODE appears.
Enter the release code and confirm by pressing ENTER; CONTROL re-appears; the setting
to be adjusted now flashes.
Use the UP or DOWN button to set to OFF and confirm by pressing ENTER. The message
appears "Data has been saved".
Set the radiation output in accordance with 6.3.2 to the maximum possible value.
Calibrate the sensor in accordance with 6.3.1.
Use the STOP/START button to switch to SYSTEM OFF.
Press the CHANGE button to go to the CONTROL display.
Confirm by pressing ENTER, REQUEST RELEASE CODE appears.
Reference value: as a % of the maximum radiation output
max
F
maximum flow rate anticipated at the flow
meter, to which the 20mA signal needs to
be
F
maximum permitted flow rate for the
max
Dulcodes M according to the data sheet
depending on UV transmission and
required dose of radiation
allocated
®