Replacing The Fan Filter Mats And Air Exit Filter; Troubleshooting - ProMinent Dulcodes M Operating Instructions Manual

Uv-system
Table of Contents

Advertisement

Maintenance / Troubleshooting
↓ ↓ ↓ ↓ ↓
55 %
10 h
Operation
20
On/Off
↓↓
↓↓
↓↓
↓↓
↓↓
40 %
Free flushing period
Time 00:11
Fault
UV Sensor
Possible cause
Remedy
Possible cause
Remedy
Possible cause
Remedy
Possible cause
Remedy
Possible cause
Remedy
Page 36
7.5

Replacing the Fan Filter Mats and Air Exit Filter

DANGER
If the filter mats in the fan or the air exit filter become clogged, the control cabinet may
overheat and be destroyed.
The fan filter mats and air exit filter should be replaced at least once a year. They must
be replaced more frequently under unfavourable environmental conditions.
Press the START/STOP key to switch off the system.
Switch off at the mains.
Remove the fan cover. Take hold of the recessed grips at the bottom of the cover and
remove it.
Remove the clogged filter mats and insert new ones with the white side facing inwards
(control cabinet side).
Replace the air exit filter mat as described above.
Switch on at the mains.
8

Troubleshooting

IMPORTANT
Troubleshooting with the control cabinet open and replacement of components may
only be carried out by an authorized skilled electrician.
Value fallen below warning threshold
Message: flashing arrow pointing downwards
Value fallen below safety threshold
Message: flashing double arrow pointing downwards The amount of free flushing
time that remains is displayed
(Instead of the seconds, two squares flash)
Value fallen below safety threshold
Message: UV sensor (when the maximum free flushing period has elapsed)
Press the ENTER key to acknowledge the fault message
Insufficient lamp output.
Increase lamp output
A coating has formed on the protective tube and/or UVC sensor.
Clean the protective tube and UVC sensor
The UV transmission of the water to be disinfected has deteriorated.
Improve the water quality
The UV lamp is at the end of its useful life.
Insert a new UV lamp
Sensor not calibrated.
Calibrate the sensor
®