Electromagnetic Compliance (Emc); Emc - Manufacturer´s Declaration - Invacare Rea Dahlia 30 User Manual

Manual wheelchair passive
Hide thumbs Also See for Rea Dahlia 30:
Table of Contents

Advertisement

Rea® Dahlia® 30 / 45
Degree of protection
For full details contact your Invacare authorized provider.

11.9 Electromagnetic compliance (EMC)

General information
Products with electronic equipment needs to be installed and used according to the EMC information in this user manual.
WARNING!
Portable and mobile communications equipment can affect the operation of this product.
This product must not move unintentionally while being submitted to Electromagnetic interference
– If this should happen; the product must immediately be taken out of use and checked by a qualified technician.
This product has been tested and complies with the EMC limits specified by IEC 60601-1-2 for Class B equipment.
This product has a very low emission and should not interfere with other equipment.
However, if other devices nearby should react inexplicably, run and stop this product and observe the devices.
If nothing happens with the other devices, then this product is not causing the error.
If other devices are behaving in an inexplicable manner, then this product is causing the error. Solve the problem
by moving or increasing the distance between them.
11.10 EMC - Manufacturer´s declaration
Test
Mains PortConducted Emissions
Radiated Emissions
Mains Harmonics
Voltage Fluctuations
Electrostratic Discharge
Radiated Field Immunity
EFT/Bursts
Surge Immunity
Conducted RF Immunity
Magnetic Field Immunity
Dips & Interruptions
52
Applied Part complying with the specified requirements for protection against electrical shock
according to IEC60601-1.
The control unit, external power supply, motors and hand controls are protected according to
IPx6. See label and label on each electric device for correct IP class. The lowest IP-classification
decides the overall classification of the system.
IPx6- The system is protected against water projected from any direction (not high pressure).
Basic Standard
EN 55011
EN 55011
EN 61000-3-2 +A1 + A2
EN 61000-3-3
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
EN 61000-4-8
EN 61000-4-11
Compliance Level / Frequency Range
Group 1 Class B
Group 1 Class B
Class A
dc%/dmax/d(t)/Pst
±2/4/6kV Contact, ±2/4/8kV Air
3V/m, 80% 1kHz AM, 80MHz - 2.5 GHz
±2kV L/N/E/LNE
±0.5/1kV L-N, ±0.5/1/2kV L-E/N-E
3V, 80% 1kHz AM, 0.15MHz - 80MHz
3A/m, 50Hz-80Hz
-95%/-60%/-30%