Actuator: Maintenance; Hsm: Maintenance - MTS Systems 318 Series Product Information

Load unit
Table of Contents

Advertisement

Routine Maintenance Overview Checklist

244 Actuator: Maintenance

Weekly
Monthly
Yearly

298 HSM: Maintenance

98
Maintenance
The 244 Actuator is designed for extended periods of operation without extensive
maintenance requirements.
A summary of the routine maintenance procedures is listed below:
Clean exposed areas of the actuator piston rod with a clean, dry, lint free rag. If
the actuator is continually exposed to a dirty operating environment, clean the
piston rod on a daily basis.
Inspect actuator piston rod and seals for excessive wear and/or leakage. Small
scratches in the axial direction of the piston rod or polishing of the rod surface is
considered normal operating wear.
Change actuator seals if necessary. Actuator assemblies can require more or less
frequent seal changes depending on usage. External oil leakage and/or decreased
performance are indicators of seal wear. Seal replacement is considered a service
procedure; contact you MTS Service Engineer for additional information.
The 298 Hydraulic Service Manifold requires the filter be changed periodically.
Dirty
Filter
Indicator
Inlet
Filter
Pressure
Accumulator
Component Identification
Return
Accumulator
318 Load Unit