MTS Systems 318 Series Product Information page 82

Load unit
Table of Contents

Advertisement

Routine Maintenance Overview Checklist
82
Maintenance
The crosshead can slowly drift down the columns if the locks are turned off
and the hydraulic pressure is turned off.
The crosshead can damage any test fixtures, grips, and specimen in its
path.
Unlock the crosshead only to reposition it. Always lock the crosshead after you
have repositioned it, and never leave the crosshead unlocked.
1. Make sure your crosshead is locked.
2. Turn on system electrical power.
3. Reset any active interlocks at the test controller.
4. Turn on high hydraulic pressure.
5. If there is a specimen in the load unit, remove it.
The lifts contain hydraulic fluid under high pressure.
If the bleed port screw is unscrewed all the way, the screw can fly out of its
port at high velocity and it could hurt you or damage your equipment.
Unscrew the bleed port screw no more than 1/2 turn to vent the trapped air.
6. Use a 1/8 inch hex key (or slotted screwdriver, depending on model number)
to open one of the bleed ports. Do not unscrew the bleed port screw more
than 1/2 turn.
CAUTION
WARNING
1/2 turn maximum
Open to Bleed
318 Load Unit

Hide quick links: