Fcb100, Fch100, Fcb400 And Fch400 Sensors In Integral And Remote Mount Design With Tower Extension - ABB CoriolisMaster FCB100 Instruction Manual

Mass flowmeter
Hide thumbs Also See for CoriolisMaster FCB100:
Table of Contents

Advertisement

Available languages

Available languages

10 – EN
CORIOLISMASTER FCX100, FCX400 SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
3.4 FCB100, FCH100, FCB400 and FCH400 sensors in
integral and remote mount design with tower
extension
NOTICE
Before the device is sealed, any settings must be
completed and the write protection switch must be set to
"write protection active" (see chapter 'FCB400, FCH400 in
integral mount design with dual-compartment housing.' on
page 5 ).
Scope of delivery
1
2
3
Fig. 15
Item
Designation
1
Holding rod
2
Locking cylinder
3
Head hole screws
Seal
Seal wire
Preparation
Fig. 16: Example of FCX100 and FCX400 sensor in remote mount
design
1.
Loosen two of the four screws
replace with head hole screws
2.
Perform any other preparatory work as described in
chapters 'FCB400, FCH400 in integral mount design
with single-compartment housing.' on page 7 , 'FCB100,
G12350
FCH100, FCB400 and FCH400 sensors in remote mount
design ' on page 9 and 'Activating the hardware write
protection' on page 3 .
Quantity
2
2
4
1
1
1
2
2x
G12364
2
on the foot and
.
1

Advertisement

Chapters

Table of Contents
loading

Table of Contents