CORIOLISMASTER FCX100, FCX400 SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
2.1

Dual- compartment housing

1
Fig. 1:
Position of the DIP switches
1 DIP switch, NAMUR 2 DIP switch, write protection
DIP switches are located behind the front housing cover.
The DIP switches are used to configure specific hardware
functions. The power supply to the transmitter must be
briefly interrupted in order for the modified setting to take
effect.
Write protection switch
When write protection is activated, device parameterization
cannot be changed via the LCD indicator. Activating and
sealing the write protection switch protects the device
against tampering.
Number
Function
On
Write protection activated
Off
Write protection deactivated.
2.1.1

Single-compartment housing

2
G11779
Fig. 2:
1 DIP switch, write protection
The DIP switch is used to configure specific hardware
functions. The power supply to the transmitter must be
briefly interrupted or an device reset have to be performed
in order for the modified setting to take effect.
Write-protect switch
When write protection is activated, device parameterization
cannot be changed via the LCD indicator. Activating and
sealing the write protection switch protects the device
against tampering
Position
On
Off
Position of DIP switch
Function
Write protection active
Write protection deactivated.