CORIOLISMASTER FCX100, FCX400 SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
3.2
FCB400, FCH400 in integral mount design with
single-compartment housing.
NOTICE
Before the device is sealed, any settings must be
completed and the write protection switch must be set to
"write protection active" (see chapter 'Activating the
hardware write protection' on page 3 ).
Scope of delivery
1
Fig. 7
Item
Designation
1
Head hole screws
Seal
Seal wire
Preparation
Fig. 8
1.
Loosen two of the four screws
replace with head hole screws
G12355
Quantity
2
1
2
2x
1
1
G12356
2
on the foot and
1
.
Lay seal wire
1
2
Fig. 9
1.
Insert the seal wire
1
housing and cover on the left side such that both ends
are the same length.
2.
Tightly twist the seal wire
eyelets
such that the twist loops are smaller than 1
2
cm.
1
2
Fig. 10:
1 Seal wire 2 Eyelets
3.
Continue to twist the seal wire
eyelets
2
and secure as previously described.
through the eyelets
2
on the
right behind the
1
up to the right
1
EN – 7
G12357
G12358