14 TROUBLE SHOOTING
Trouble
No indication
Weight indication is
changing permanently
Weighing result is obviously
false
14.1 Error messages
O-Err
The balance has been overloaded.
b-Err
Electronic error, caused by electrostatic fields. Check the
environment. When necessary change the installation place.
Wrong adjusting weight (too light).
1-Err
Wrong adjusting weight (to heavy).
2-Err
Transport lock not released, see point 3 "Installation". Weighing
11-Err
mechanism defective. Please contact the manufacturer.
When other error messages appear turn off the balance and on again. When the error
message still appears please contact the manufacturer.
470-BA-defs-0017
Possible causes
Balance is not turned on
-
Mains connection does not function (cable not connected,
-
cable defective).
Mains supply break down.
-
-
Air movements.
-
Vibrations of the table or the floor
-
The pan is in contact with an alien element.
-
Balance has not been tared.
-
No adjusting has been done.
-
Temperature deviations.
59