Trouble Shooting - KERN MCN-M Service Manual

Medical chair scale
Table of Contents

Advertisement

6.5 Trouble Shooting

Power on
Full Segments?
If no display,
• check battery / power adaptor,
• check connection between keyboard—main board,
• check connection between battery / power adaptor—main board.
If missing segments,
• check fixing of LCD frame, zebra connector under LCD.
Display Zero?
If display [L0 bat], check battery >6.5V, power adaptor ~6V.
If display [ zerohi ] or [ zerolo ] check the load cell connections,
load cell damaged?
If display [ instab ] check weighing platform, overload stopper, load cell and wires,
environment and stable floor.
Proper readout?
If unstable reading, check weighing platform, overload stoppers, loadcell and wires,
environment and stable floor.
If always zero, check ADC. Linearization if necessary.
Correct reading?
If not accurate, perform adjustment or linearization.
If cannot reach full capacity, check weighing platform, overload stopper, load cell.
Buttons functioning?
If some buttons not working, check connection wires. Replace keyboard if needed.
Interface functioning?
If interface not working, check the connection wires.
Normal operation.