Fault – Cause – Remedy
Locating and remedying leakages
Leakages can lead to a failure of the system test or
the leakage test and must be remedied.
CAUTION
Risk due to contamination
Anesthetic gas can get into the ambient air as a
result of leakages.
Perform the leakage test before using the device.
Remedy all leakages.
CAUTION
Risk of insufficient ventilation
Breathing gas may escape because of leakages,
with the result that the applied volume is less than
the set volume.
Perform the leakage test before using the device.
Remedy all leakages.
Cleaning
Disassembly
Sequence of disassembly
1 Remove the breathing bag.
2 Loosen the knurled screws.
3 Unscrew the rigid arm from the compact breath-
ing system.
Supplement Fabius
Possible causes of leakage
– The breathing bag is improperly attached or
damaged.
– The holder for the breathing bag is incorrectly
mounted to the breathing system. The sealing
ring is soiled or damaged.