English US
WARNING
Risk of too high airway pressure or insuffi-
cient fresh-gas
If the breathing bag is pinched, excessive air-
way pressures or a lack of fresh gas can oc-
cur.
Attach and position the breathing bag so that
it is not pinched and can inflate freely.
CAUTION
Risk of patient recovering consciousness
If no breathing bag is attached, a lack of fresh gas
can occur. Manual ventilation is also impossible in
this situation.
Make sure that the breathing bag is connected
tightly to the bag holder during the ventilation.
Fitting the rigid arm
Before mounting the rigid breathing bag holder,
the bag elbow on the compact breathing system
must be removed.
1 Position the connection adapter (A) of the rigid
arm on the connector of the compact breathing
system and align it.
2 Screw the knurled screws (B) tight. Ensure that
the rigid arm is attached securely!
3 Fasten the breathing bag (C) at the other end of
the rigid arm.
14
C
B
A
Supplement Fabius