Emerson Rosemount MCL-220 Reference Manual page 44

Monochloramine system with transmitter
Hide thumbs Also See for Rosemount MCL-220:
Table of Contents

Advertisement

Programming the transmitter
May 2019
6. To make a change, move the cursor to the desired line and press ENTER.
7. Press ENTER to store the setting.
8. To return to the main display, press MENU and then EXIT.
6.4.4
Simulate alarms
Complete the following steps to make the transmitter energize or de-energize an alarm
relay.
Procedure
1. Press MENU.
2. Move the cursor to Program and press ENTER.
3. Choose Alarms.
4. Choose Simulate.
44
On time
Recover time
Hold while active
The last four items describe the operation of the timer. Only four items are shown at
a time. To view the remaining items, scroll to the bottom of the screen and
continue scrolling.
A screen appears in which the present setting can be edited.
For an explanation of terms, see
The main Menu screen appears.
Purpose
and Definitions.
Reference Manual
00809-0100-3414
Emerson.com/Rosemount

Advertisement

Table of Contents
loading

Table of Contents