Emerson Rosemount MCL-220 Reference Manual page 36

Monochloramine system with transmitter
Hide thumbs Also See for Rosemount MCL-220:
Table of Contents

Advertisement

Programming the transmitter
May 2019
Ranging an
output
6.3.3
Configure outputs
Complete the following steps to configure the analog current outputs.
Procedure
1. Press MENU.
2. Move the cursor to Program and press ENTER.
3. Press ENTER.
4. Choose Configure.
5. Choose Output 1 or Output 2.
36
programmed in this section. There are two output fault modes: fixed
and live. Fixed means the selected output goes to the previously
programmed value (between 0.00 and 22.00 mA) when a fault occurs.
Live means the selected output is unaffected when the fault occurs.
The outputs are fully rangeable, including negative numbers. If the
output is logarithmic, assigned values must be positive.
The main Menu screen appears.