Reference Manual
00809-0100-3414
6. To make a change, move the cursor to the desired line and press ENTER.
7. Press ENTER to store the setting.
8. To return to the main display, press MENU and then EXIT.
6.3.4
Range outputs
Complete the following steps to range the outputs by assigning values to the low and high
outputs.
Procedure
1. Press MENU.
2. Move the cursor to Program and press ENTER.
3. Press ENTER.
4. Choose Range.
Rosemount MCL
The screen shows the present configuration. There are six items: Assign (S1 is
sensor 1), Range, Scale, Dampening, Fault Mode, and Fault Value. To display the
fifth and sixth items, scroll to the bottom of the screen and continue scrolling.
A screen appears in which the present setting can be edited.
For an explanation of terms, see
The main Menu screen appears.
The cursor is on Outputs.
Programming the transmitter
Purpose of configuration
May 2019
and Definitions.