CALIBRATION
Occasionally the scale should be verified that it is weighing correctly by measuring a
known mass of approximately the scale capacity. Tare the scale, place the mass on the
centre of the platform and note the reading. Calibrate the scale if necessary.
Before beginning make sure you have the mass specified and you know the mass is
accurate. The scale can calibrate using either kilograms or pounds. See the below for
the mass required. The following procedure assumes the parameter for local gravity
correction "Grav" has been set to OFF. If it has been set to ON then reset the value to off
as shown in the next section.
PROCEDURE
Turn the scale off.
Press and hold the On/Off key then press the Units key 3 times. Release the On/Off key.
The display will show a value from the internal counters. The typical value is 2000 to 8000.
Press the Tare key. Display will show zero.
Press the Unit key to select either kilograms (kg) or pounds (lb).
Place the calibration weight on the scale, centred on the platform.
CPW 25
CPW 50
CPW 100
CPW 150
The display will show a value from the internal counters again.
Press the Tare key again.
The display will show " CAL" then return to weighing showing the value of the weight
used.
Remove the weight. Verify the scale is calibrated correctly.

SET FUNCTIONS

The scale can be set as desired for the user. The settings that can be changed are:
Weighing Units enabled
Time delay before the scale automatically turns off.
Enable or disable the correction for the local gravity.
Press and hold the TARE key for 3 seconds until the display shows "Set" then the first
menu option "Unit"
© Adam Equipment Company 2002
20kg
OR
40kg
OR
100kg
OR
100kg
OR
5
50 lb
100 lb
200 lb
200 lb