CAUTION!
9742146
Note to the customer
The multi-washer and disinfection machine is provided for the permanent connection to
the electrical power supply and the connection to the on-site potential equalisation
system and is accordingly equipped with a connection option.
The operator can decide at his/her own discretion and under his/her own responsibility to
alternatively implement personal protection in collaboration with a licensed electrician
company using:
• Fault current protection switch sensitive to universal current with max. 30 mA EN
62423
or
• Automatic shut-down of the supply when the protective earth conductor for the
consistency is lost (EN 60204-1 Chap. 8.2.8.c)
5.5

Requirements for the fresh water connection

The appliance is DVGW-compliant and does not require an extra safety valve in the
water supply.
The fresh water connection must be effected as per EN 1717 or local regulations.
The appliance is equipped with an open discharge section (group A, model A as
per DIN EN 1717).
The requirements for the clean water supply in the accompanying GiO module
operating and service manual must be observed for appliances equipped with the
GiO module.
The minimum flow pressure of the fresh water supply must be at least 0.6 bar, and
in appliances with a GiO module, there must be at least 1 bar before the solenoid
valve.
The maximum pressure must not exceed 5 bar.
If the minimum flow pressure is not reached, increase the flow pressure with a
booster pump; if the maximum pressure is exceeded, limit it with a pressure
regulator.
A water stop is integrated into the appliance's fresh water line. This, together with
the leakage detector in the floor pan of the base, ensures that the clean water
supply is shut off in the event of a leak.
Suitable protective measures must be taken to ensure that no iron particles can
enter the appliance via the mains water supply. Similarly, precautions must be
taken to prevent the entry of other metal particles, such as copper turnings.
Corresponding instructions are contained in the installation drawing. Appropriate
measures must be taken.
A dirt trap must be fitted into the clean water supply to protect the solenoid valve.
5.6

Requirements for the waste water connection

A waste water pump is integrated into the drain pipe.
The drain hose must be connected to the locally available waste water pipe.
The requirements for the waste water supply in the accompanying GiO module
operating and service manual must be observed for appliances equipped with the
GiO module.
5.7

Emergency off

Disconnect the appliance from the power supply using the locally available main
switch.